Chapter G - PARTS REPLACEMENT & REPAIR
G-12 Planmeca Compact i
INSTRUMENT MULTIPLEXER RELATED
Technical Manual
3.4 Replacing multiplexer control membrane
Every multiplexer block, except the syringe (next to the control panel) has a control mem-
brane, which blocks air and water for the instrument (instrument not active).
NOTE
Care must be taken to position the new multiplexer control membrane cor-
rectly.
NOTE
Carefully tighten the screws equally and crosswise when repositioning the
multiplexer block cover. Using too much force might damage the threads in the
multiplexer block.
a)
Perform the pre-replacement preparations described in section 3.1 “Preparations before
instrument multiplexer related replacements” on page G-7.
b)
Unscrew the four TORX WN1451 3x22 PT screws and one TORX WN1451 3x12 PT screw
from the multiplexer block to be repaired and remove the block cover.
c)
Remove the spring and membrane press.
d)
Install the new control membrane in reverse order.
TORX WN1451 3x12 PT
TORX WN1451 3x22 PT
Block cover
Multiplexer control
membrane
Spring
Membrane press
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......