Chapter D - HELP & ERROR MESSAGES
D-4 Planmeca Compact i
HELP MESSAGES
Technical Manual
HE 27
Compact WE’S help message, should not
appear in Compact i: The chair seat cannot be
driven down, the backrest is too low.
Drive the backrest upwards before lowering
the seat. If this help message is shown while
the patient chair is attached to the unit, it
means that you have forgotten to switch on the
chair’s automatic positions. To clear the mes-
sage, put the operation mode switch on the
cuspidor side of the chair base in the “AUTO”
position (switch up).
HE 28
Compact WE’S help message, should not
appear in Compact i: The patient chair cannot
be driven up/down because the lifter cable is
not connected.
Connect the lifter cable.
HE 29
Compact WE’S help message, should not
appear in Compact i: The chair backrest can-
not be driven down, the seat is too high.
Drive the seat downwards before lowering the
backrest. If this help message is shown while
the patient chair is attached to the unit, it
means that you have forgotten to switch on the
chair’s automatic positions. To clear the mes-
sage, put the operation mode switch on the
cuspidor side of the chair base in the “AUTO”
position (switch up).
HE 30
The unit is waiting for the container to be
attached.
Attach the container (or make the long rinse
cycle or use service mode “n.101”).
HE 31
The unit’s power can now be turned off.
Turn off the unit if needed. When the unit is
turned back on, remove the container, fill it
with water and attach it again to the unit (or
make long rinse cycle or use service mode
“n.101”).
HE 32
The pre-rinsing cycle is completed and the
container should be removed.
Remove the container (or make long rinse
cycle or use service mode “n.101”).
HE 33
The unit is waiting for the main water to be
turned on.
Remove the container and open the tap (or
make long rinse cycle or use service mode
“n.101”).
HE 34
No instruments were selected when starting
the waterline cleaning operation.
Select (remove from holder, twist doriot) at
least one instrument (not syringe).
HE 35
The instruments are locked for safety reasons. Press the instrument spray key to perform the
instrument flushing.
HE 36
After flushing (short, long or Waterline Clean-
ing System): Remove the syringe from the
flushing holder and return it back. If the
syringe is already removed, return it back to
the flushing holder. This ensures that no water
will flow from the syringes after flushing.
HE 37
The Waterline Cleaning System cycle has
been interrupted.
HE 38
Waterline cleaning system help message: Low
water pressure during the final rinsing.
HE 39
Waterline cleaning system help message:
Make sure that the syringe is properly posi-
tioned in the flushing holder before the WCS
procedure.
The help message will disappear after bend-
ing the syringe instrument arm and returning it
back to its holder.
Help Code
Meaning
Action or Explanation
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......