Chapter F - ADJUSTMENTS
Planmeca Compact i F-37
WMS OPERATION
Technical Manual
Cleaning agent is fed to internal water lines
City Water
Bottled Water
Feeding cycle:
Internal water lines
Similar to City water mode
V1
c
V2
o
V3
o
V4
o
V5
c
Pump
running
Display
countdown timer 30 s
Duration
30 s
Excess cleaning agent is removed from bottle
City Water
Bottled Water
Feeding cycle:
Emptying bottle
Similar to City water mode
V1
c
V2
c
V3
o
V4
o
V5
c
Pump
not running
Display
wait, when L2 reached countdown
timer 15 s
Duration
until L2 r 15 s
When feeding cycle is completed, Display: He 66
Flushing cycle is started
City Water
Bottled Water
After 8 hours if unit let on or when unit restarted
on
When unit turned on
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......