Chapter F - ADJUSTMENTS
Planmeca Compact i F-27
PNEUMATICAL ADJUSTMENTS
Technical Manual
3.2 Adjusting syringe water and air flow rates
The water and air flow rates for the syringe are adjusted from the two brass screws located on
the syringe instrument multiplexer block. Both the water and air flows for the syringe can also
be blocked by these screws.
a)
Open the instrument console cover by unscrewing the M6x16 DIN 7984 screw from under the
instrument console and open the console cover.
b)
Adjust the syringe water flow rate from the water flow rate adjuster on the top of the syringe
instrument multiplexer block. Observe the adjusted flow rate with the syringe instrument.
c)
Adjust the syringe air flow rate from the air flow rate adjuster on the instrument arm side of the
syringe instrument multiplexer block. Observe the adjusted flow rate with the syringe
instrument.
3.3 Adjusting coolants to the instruments
The water and air flow rates for each instrument are adjusted in the instrument settings pro-
gram mode. Please refer to the User’s manual (publication part number 288037) for program-
ming the instrument settings.
Water flow rate adjuster
Air flow rate adjuster
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......