Chapter F - ADJUSTMENTS
Planmeca Compact i F-35
WMS OPERATION
Technical Manual
6.3 WMS operation, WCS use
When door is open, pump is not running and syringe cannot be used.
City Water
Bottled Water
Door open
Door open
V1
c
V1
c
V2
c
V2
c
V5
c
V5
c
V6
c
V6
c
Pump
not running
Pump
not running
Display
door
Display
door
Syringe
no water flow
Syringe
no water flow
When suction is used, the main water valve is open to feed water to suction system.
City Water
Bottled Water
Suction in use
Suction in use
V1
o
V1
o
V4
o
for 150 ms (from software version
3.07)
V4
o
for 150 ms (from software version
3.07)
When bottle is removed water cannot be used.
City Water
Bottled Water
Special situation: bottle removed
Special situation: bottle removed
SW2
off
SW2
off
V3
c
V3
c
V2
c
V2
c
V5
c
V5
c
Pump
not running
Pump
not running
Display
He 62
Display
He 62
Continuously
City Water
Bottled Water
SW2
on
SW2
on
V6
c
V6
c
V7
o
V7
o
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......