Chapter C - SERVICE MODE
Planmeca Compact i C-31
SERVICE MODE DETAILED DESCRIPTION
Technical Manual
In mode 3 the instruments are controlled with horizontal movement of the pedal, and the
reversing of the micromotor rotation is done with foot control as follows. To operate the
micromotor normally (in a clockwise direction) push the pedal to the right. To reverse the
direction of rotation push the pedal to the left. Note, that the reverse rotation of micromotor
key does not function in this mode.
In mode 4 the Plus package power limitation selection is done with foot control with all
instruments.
n.67
Disable foot control pedal extra functions (n mode = 0, 1)
This service mode can be used to disable the foot control pedal extra functions. Number 0
(zero) removes ALL extra features. Then the foot control pedal can be used only to control the
instruments. Number 1 (one, factory default) enables all possible pedal functions. This cur-
rently includes; EXT- function control (nurse call/door open/table instrument control), cup fill-
ing + cuspidor rinsing, and patient chair S position control. Other mode numbers are reserved
for future use.
n.68
Disable automatic chair positions (A, B, C, D) from foot
control (n mode = 0, 1, 2)
This service mode can be used to disable the foot controlled pre-programmed chair positions.
Number 0 disables these automatic functions from the chair control knob. Then this knob only
operates the chair manually (backrest up/down, seat up/down). Number 1 (factory default)
enables the automatic chair position control (short activation of the knob runs chair to a pre-
programmed position). Number 2 disables both automatic and manual chair controls.
n.69
Momentary spray mode / spray with foot control mode (n
mode = 0, 1, 2, 3, 4, 5, 6, 7)
This service mode can be used to select the momentary spray and spray with foot control
modes. The momentary spray can either be
totally
disabled or momentary spray water can
be switched off. The instrument spray operated with foot control
can also be disabled in this
service mode, when the spray can only be operated from the control panel
.
NOTE
The momentary spray is activated when the pedal is pressed down while
instrument in use.
0 = momentary spray
water off
, instrument spray with foot control
on
(default)
1 = momentary spray
water on
, instrument spray with foot control
on
2 = momentary spray
water off
, instrument spray with foot control
disabled
3 = momentary spray
water on
, instrument spray with foot control
disabled
4 = momentary spray
disabled
, instrument spray with foot control
on
5 = factory use
6 = momentary spray
disabled
, instrument spray with foot control
disabled
7 = factory use
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......