Chapter F - ADJUSTMENTS
Planmeca Compact i F-33
WMS OPERATION
Technical Manual
Cup fill is used, no need to fill bottle.
City Water
Bottled Water
Cup fill (water bottle full)
Cup fill (water bottle full)
V1
c
V1
c
V2
c
V2
c
V5
o
V5
o
V6
c
V6
c
Pump
not running
Pump
not running
Display
PCI
Display
PCI
Cup fill is used and at same time bottle is filled.
City Water
Bottled Water
Cup fill (water level below upper sensor)
Cup fill (water level below upper sensor)
V1
o
V1
c
V2
o
V2
c
V5
o
V5
o
V6
c
V6
c
Pump
running
Pump
not running
Display
PCI
Display
PCI
Cup fill is used and at the same time bottle is filled. In city water mode automatic cup fill does
not work.
Automatic cup fill = cup fill key pressed once
Manual cup fill = cup fill key pressed continuously
City Water
Bottled Water
Cup fill (water level below lower sensor)
Cup fill (water level below lower sensor)
Automatic cup fill
Manual cup fill
Automatic cup fill
Manual cup fill
V1
o
o
V1
c
c
V2
o
o
V2
c
c
V5
c
o
V5
o
o
V6
c
c
V6
c
c
Pump
running
running
Pump
not running
not running
Display
He 60
He 60
Display
He 61
He 61
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......