Chapter G - PARTS REPLACEMENT & REPAIR
Planmeca Compact i G-35
MOTORS REPLACEMENT
Technical Manual
k)
Unscrew the eight M6x45 DIN 912 screws that hold the nut attachment plate in position.
l)
Remove the nut attachment plate and the lifting frames.
m)
Remove the ø10 DIN 6799 locking ring from the lift motor attachment axle.
n)
Slide the lift motor from the attachment axle and turn it in a way that the Motor grounding
cable can be detached. Detach the Motor grounding cable from the side of the motor.
o)
Lift the Lift motor from the cuspidor.
p)
Unscrew the M10x25 ULS screw from the end of the worm screw and remove the worm
screw nut.
q)
Detach the lift motor position sensor by unscrewing the two M5x12 DIN 7500 screws.
r)
Grease the new motor’s worm screw. Install the new lift motor in reverse order.
NOTE
Pay attention when reconnecting the lift motor position sensor, refer to section
5.8 “Replacing position sensors” on page G-26.
NOTE
Calibrate the lift motor position sensor as described in section 2.2 “Calibrating
lift motor position sensor” on page F-20.
Unscrew the eight screws
Ensure that the steel bar
is supporting the whole
weight of the chair!
Remove the nut attachment plate
and the lifting frames
Remove the locking ring
Detach the grounding cable
Electronics
control box
lifted aside
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......