output of relay NO 4 on Phoenix Contact relay is 5VDC. Check the connection on
DPM carrier board and the output on DPM carrier board for the blower. Check the
Blower Speed Controller output.
F
IGURE
60
-
BLOWER
STATUS
P
AGE
12.7 Ambient Temperature
From the SYSTEM OVERVIEW page, clicking on the A. TEMP box will take the user to
the AMBIENT TEMP STATUS page (Figure 61). This will allow operators to monitor the
current temperature of the DPM reader in °F and °C. The temperature is measured from
the inside of the enclosure wall located close to the Humidity Filter. The temperature
reported by the sensor is normally 5°C above the external ambient temperature of the
DPM Reader.