1. Turn on DPM Reader.
2. Test live (brown) and neutral (Blue) input cables of PSU. Voltage should read
mains voltage 110-240VAC.
3. Test DC output voltage from PSU. Orange Cable is marked with a '+' on the
PSU board and ground (black) is the alternative pin. The voltage should be
above 14.5VDC.
F
IGURE
23
-
AC
AND
DC
V
OLTAGE
T
EST
P
OINTS IN THE
PSU
9.1.6.3
80358
P
OWER
S
UPPLY
U
NIT
R
EPLACEMENT
Removal
Warning: Set DPM Reader to Safety Control Level 4
Refer to Figure 24.
1. Unplug 3 connectors
2. Using an M5.5 socket remove the 4 nyloc nuts and washers (do not re-use the
nyloc nuts).