PIN-M-02 DPMRA01.A Service Manual
Copyright © Pinssar (Aust) Pty Ltd 2019. The PINSSAR logo is a registered trademark of Pinssar Pty Ltd
Page 46 of
96
9.1.4
112204 Battery Replacement
Removal
Warning: Set DPM Reader to Safety Control Level 4
1. Unplug 2 connectors.
2. Using a No.2 Phillips screwdriver, remove the 2 screws from the battery clamp.
3. Carefully remove battery, ensuring battery terminals do not touch any metal
parts.
F
IGURE
21
-
B
ATTERY
R
EMOVAL
Installation
Warning: Set DPM Reader to Safety Control Level 4
1. Place clamp over the battery and fix with 2 screws using a No.2 Phillips head
screwdriver.
Reminder: Ensure the red wire is connected to the positive terminal and the
black wire to the negative terminal which are clearly marked on the battery.
Ensure a tight fitting of the cable terminals onto the corresponding fixed
battery terminals.
2. Allow the battery to charge for at least 4 hours in SCL 1C before turning the unit
on.
3. With the unit in SCL 4, open circuit test the battery. Voltage should be greater
than 12.8V.
9.1.5
Isolation Switch Fault Servicing
9.1.5.1
T
EST
S
WITCH
F
UNCTION
Warning: Set DPM Reader to Safety Control Level 1A
Step 1
Step 2