WARNING
Inspect all components and the transducer. Ensure that the biopsy guide you are using is the
correct one for the transducer, the system, and system software. Your Philips representative
can verify this information for you.
WARNING
Use only Philips-approved biopsy guides, transducers, covers, brackets, supplies,
components, and accessories. Other brands may not properly fit Philips transducers.
Improper installation may result in patient injury.
WARNING
Some biopsy guides must be installed over a sterile transducer cover.
WARNING
After each use, biopsy guides must be either sterilized or disposed of, depending upon the
type. See the instructions included with the biopsy guide.
WARNING
Most transducers can only be disinfected; they cannot be sterilized. Only the transducer
cover provides the sterile barrier.
Biopsy Guideline Display
WARNING
Do not attempt to use the biopsy guide until you have read the instructions for selecting the
display, installing the sterile transducer cover, and verifying alignment of the biopsy guide.
Biopsy Guides
Biopsy Guideline Display
246
EPIQ 7 User Manual 4535 617 25341
Summary of Contents for epiq 7
Page 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Page 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Page 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Page 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Page 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Page 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...