For information on allergic reactions to latex-containing medical devices, see
Transmissible Spongiform Encephalopathy
WARNING
If a sterile transducer cover becomes compromised during an intraoperative application
involving a patient with transmissible spongiform encephalopathy, such as Creutzfeldt-Jakob
disease, follow the guidelines of the U.S. Centers for Disease Control and this document from
the World Health Organization: WHO/CDS/ APH/2000/3, WHO Infection Control Guidelines
for Transmissible Spongiform Encephalopathies. The transducers for your system cannot be
decontaminated using a heat process.
Choosing the Correct Transducer-Care Method
To choose an appropriate care method for your transducer, you first must determine the
transducer's classification, based on its use.
Care Methods by Transducer Type
Transducer Use
Example
Classification
Care Method
Contacts intact skin
Curved, linear, xMatrix, and
sector transducers
Noncritical
Low-level disinfection (see
“Transducer, Cable, and Connector
Cleaning” on page 258
Contacts mucous
membranes
Endocavity and
transesophageal (TEE)
Semi-critical
High-level disinfection (see
“Transducer, Cable, and Connector
Cleaning” on page 258
“Transducer Disinfection” on page
263
Enters otherwise
sterile tissue
Intraoperative
Critical
and Connector Cleaning” on page 258
and
“Transducer Sterilization” on page
Transducer Care Safety
Transducer Care
EPIQ 7 User Manual 4535 617 25341
257
Summary of Contents for epiq 7
Page 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Page 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Page 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Page 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Page 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Page 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...