The care method for your transducer determines the appropriate disinfectant for your
transducer. For details about compatible disinfectants, see
“Disinfectants Compatibility” on
Transducer, Cable, and Connector Cleaning
These general cleaning (low-level disinfection) instructions are indicated for all transducers and
cables. It is important that you clean the transducer and cable according to the following
procedures.
WARNING
Always use protective eyewear and gloves when cleaning and disinfecting any equipment.
WARNING
If a premixed solution is used, be sure to observe the solution expiration date.
WARNING
Transducers must be cleaned after each use. Cleaning the transducer is an essential step
before effective disinfection or sterilization. Be sure to follow the manufacturer’s instructions
when using disinfectants.
CAUTION
Attempting to disinfect a cable or connector by using a method other than the one included
here can damage the device and voids the warranty.
CAUTION
Do not allow sharp objects, such as scalpels or cauterizing knives, to touch transducers or
cables.
Transducer Care
Transducer, Cable, and Connector Cleaning
258
EPIQ 7 User Manual 4535 617 25341
Summary of Contents for epiq 7
Page 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Page 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Page 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Page 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Page 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Page 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...