This section includes information on electromagnetic emissions and immunity as it applies to
the system. Ensure that the operating environment of your system meets the conditions
specified in the referenced information. Operating the system in an environment that does not
meet these conditions may degrade system performance.
The information and warnings contained in this and other sections should be observed when
installing and using the system to ensure its EMC.
NOTE
See the other electrical-safety warnings and cautions in this section.
If the system is operated within the electromagnetic environment described in
“Electromagnetic Immunity” on page 74
, the system will remain safe and will provide the
following essential performance:
• Imaging
• Doppler audio and spectral display
• Measurements
• Acoustic output
• ECG triggering
• Recording and playback on a VCR or DVD recorder
• Printing using system printers
• Patient information
• Date and time information
Radio-Frequency Emissions
The following information applies to the system and any radio-frequency device included in or
with the system. For information on related labeling, see
.
FCC and Industry Canada Radio Compliance: This device complies with Part 15 of the FCC Rules
and RSS-210 of Industry Canada. Operation is subject to the following two conditions:
• This device may not cause harmful interference.
• This device must accept any interference received, including interference that may cause
undesired operation. Any changes or modifications to this equipment not expressly
approved by Philips may cause harmful radio frequency interference and void your
authority to operate this equipment.
Electromagnetic Compatibility
Safety
EPIQ 7 User Manual 4535 617 25341
69
Summary of Contents for epiq 7
Page 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Page 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Page 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Page 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Page 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Page 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...