• Check the patient’s history for gastroesophageal disease or difficulty swallowing.
• Evaluate the potential overall effects of any treatment that the patient is undergoing, such
as mediastinal radiation, chemotherapy, anticoagulation, or steroid therapy.
• Be aware that you may discover unsuspected esophageal pathology during a study. Be alert
for congenital problems with the esophagus or stomach, particularly with pediatric patients.
• When examining a patient with an above-normal temperature, use the Auto-Cool feature
and enter the patient temperature. The Auto-Cool feature is described in
.
This list is not comprehensive. Rather, it suggests areas to investigate when considering TEE for
a particular patient.
Patient Selection for TEE Transducer Use
Although the TEE transducers can provide clinical data not available from other instruments,
you should consider which patients can safely use the TEE transducers.
WARNING
The ability of a patient to swallow or accommodate the transducer should be considered.
WARNING
You must consider any history of gastroesophageal diseases as well as the possible effects of
other therapies that the patient is undergoing. You must also consider all gastroesophageal
abnormalities or difficulty swallowing.
The following table lists the minimum patient weight recommendations when using TEE
transducers.
TEE Transducer Minimum Patient Weight Recommendations
TEE Transducer
Minimum Patient Weight
S7-3t
3.5 kg (7.7 lb)
X7-2t
30 kg (66 lb)
Transesophageal Transducers
Special Considerations for TEE Studies
224
EPIQ 7 User Manual 4535 617 25341
Summary of Contents for epiq 7
Page 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Page 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Page 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Page 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Page 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Page 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...