WARNING
Do not use nonmedical peripherals, such as report printers, within 1.5 m (5 ft) of a patient,
unless the nonmedical peripherals receive power from an isolated power outlet on the Philips
ultrasound system, or from an isolation transformer that meets medical safety standards, as
defined by standard IEC 60601-1.
WARNING
The system and patient-applied parts meet the standard IEC 60601-1. Applied voltages
exceeding the standard, although unlikely, may result in electrical shock to the patient or
operator.
WARNING
Connection of optional devices not supplied by Philips Ultrasound could result in electrical
shock. When such optional devices are connected to your ultrasound system, ensure that the
total system earth leakage current does not exceed 500 µA.
WARNING
To avoid risk of electrical shock, do not use any transducer that has been immersed beyond
the specified cleaning or disinfection level.
WARNING
To avoid risks of electrical shock and fire hazards, inspect the system power cord and plug
regularly. Ensure that they are not damaged in any way.
WARNING
Do not drape the power cord over any of the cable hooks or the handle on the system cart.
Damage to the cord or power receptacle unit can occur if the cart is raised.
WARNING
Operating the system with physio input signals that are below the specified minimum levels
may cause inaccurate results. See the
section.
Electrical Safety
Safety
EPIQ 7 User Manual 4535 617 25341
31
Summary of Contents for epiq 7
Page 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Page 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Page 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Page 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Page 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Page 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...