WARNING
Off-cart devices connecting to the ultrasound system must comply with the applicable IEC or
national standards, such as IEC 60601-1, IEC 60950, or the equivalent.
CAUTION
Using accessories, transducers, peripherals, or cables not supplied with the ultrasound
system or recommended by Philips can affect the system in the form of increased emissions
or decreased immunity to external EMI/EMC occurrences. Non-specified peripherals, and
cables in some cases, can also increase leakage current or compromise the safety of the
grounding scheme.
CAUTION
If systems, transducers, and peripherals have been in an environment below 10°C (50°F),
allow them to reach room temperature before connecting or turning them on. Philips
recommends allowing 24 hours for complete normalization. Otherwise, condensation inside
the devices could cause damage. If the device was only briefly exposed to temperatures
below 10°C (50°F), then the time required for the device to return to room temperature could
be significantly less than 24 hours.
NOTE
Any device that is not purchased from Philips or a Philips-authorized agent is not covered under
a Philips service agreement or warranty.
External Printers
You can connect different external printers to your system.
WARNING
Images printed on a report printer are intended only for reference and should not be used for
diagnostic purposes.
Preparing the System
Connecting Devices
102
EPIQ 7 User Manual 4535 617 25341
Summary of Contents for epiq 7
Page 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Page 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Page 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Page 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Page 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Page 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...