The TEE transducers are classified as Type BF isolated patient-applied parts, as described in IEC
60601-1. There are no exposed conductive surfaces distal to the transducer handle. To ensure
safe operation of this transducer, read the cautions and warnings in the
especially those that address electrosurgical units, pacemakers, and defibrillators.
The following table summarizes patient safety problems, describes how to prevent them, and
lists the sections in this manual where details are provided.
WARNING
If you encounter an irregularity not listed in the following table, do not use the transducer.
Potentially serious consequences could result. Contact your Philips representative.
Ensuring Patient Safety During TEE Studies
Problem
Effect on Patient
Prevention
See
Mechanical
damage
Severe trauma, cuts,
bleeding, perforations
Inspect the transducer, using both sight
and touch, before the study.
“Checking the TEE
Transducer” on page
223
Electrical
damage
Esophageal burns
Check the transducer for frayed insulation,
kinks, or other abnormalities. Follow
procedures for checking electrical safety.
“Electrical Safety and
TEE Transducers” on
page 205
Biting, scraping
transducer
Tooth damage,
esophageal burns
Always use a bite guard.
Insufficient
cleaning
protocol
Spread of illness or
disease
Thoroughly clean and disinfect the
transducer after each use. Cover the tip
and shaft with a transducer cover. Cover
the imaging system with a disposable
drape if highly pathogenic organisms are
known or suspected.
section
Transesophageal Transducers
Patient Safety During TEE Studies
202
EPIQ 7 User Manual 4535 617 25341
Summary of Contents for epiq 7
Page 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Page 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Page 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Page 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Page 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Page 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...