Rotating the S7-3t TEE Array
CAUTION
Do not use excessive force on the array rotation control wheel at its outer positions because
this may damage the endoscope mechanism.
1. To rotate the imaging array, turn the array control (the larger wheel).
2. To rotate the array to 90 degrees, align the white bar on the side of the array control with
the center of the proximal end of the transducer handle.
3. To rotate the array to a specific setting, align a setting on the top of the array control with
the center of the proximal end of the transducer handle.
Rotating the S7-3t Imaging Array
1
Array control
2
Align settings on the control with the proximal end of the transducer handle (90° setting as shown).
Transesophageal Transducers
Using the S7-3t TEE Transducer
222
EPIQ 7 User Manual 4535 617 25341