The batteries charge whenever the system is connected to AC power.
CAUTION
The batteries are used to ensure quick system startup and easy movement of the system
only. You cannot perform scans while in battery mode.
CAUTION
When shipping the system, make sure the batteries are inserted upside down to prevent
electrical discharge.
NOTE
When the system is turned on and in Sleep Mode, the batteries power the system.
NOTE
When the battery-charge level reaches a critically low state, the system shuts down
automatically.
NOTE
The batteries may need to be replaced. For information on replacing the batteries, see the
system Service Manual.
Sleep Mode
Sleep Mode uses battery power and allows quick startup of the system. It does not eliminate
power consumption and should be used only for short periods of time, such as when you
transport the system between exam locations. The Sleep Mode control (
) is in the upper left
corner of the control panel, next to the
(On/Off) control. When the system is in Sleep Mode,
it freezes live imaging, pauses any open study, exits any QLAB study, and terminates any
remote connection. Awaken the system to resume normal system operation.
Power Management
Using the System
EPIQ 7 User Manual 4535 617 25341
131
Summary of Contents for epiq 7
Page 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Page 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Page 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Page 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Page 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Page 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...