CAUTION
Avoid disinfectant contact with the connector label.
CAUTION
Be sure to use the proper concentration of enzymatic cleaner and rinse thoroughly.
CAUTION
Do not use paper products or products that are abrasive when cleaning the transducer. They
damage the soft lens of the transducer.
CAUTION
Before storing transducers, ensure that they are thoroughly dry. If it is necessary to dry the
transducer lens after cleaning, use a soft cloth and a blotting motion, instead of a wiping
motion.
Cleaning a Transducer
NOTE
All imaging transducers are rated at a minimum IPX7 in accordance with IEC529. This rating
indicates that the device is protected against the effects of immersion. Continuous wave
(nonimaging) transducers are rated IPX1. This rating indicates that the device is protected
against the effects of vertically falling water. For all TEE transducers, the control module is rated
IPX1, while the endoscope is rated IPX7, in accordance with IEC529.
Before cleaning a transducer, read the warnings and cautions in
Connector Cleaning” on page 258
“Transducer Care Safety” on page 255
1. After every patient study, use a moist cloth to remove the ultrasound transmission gel from
the transducer.
Transducer Care
Transducer, Cable, and Connector Cleaning
260
EPIQ 7 User Manual 4535 617 25341
Summary of Contents for epiq 7
Page 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Page 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Page 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Page 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Page 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Page 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...