4. Use a soft cloth lightly dampened in a mild soap or detergent solution to clean the cable,
the strain reliefs, and the connector. A soft-bristled brush can be used to clean only the
metal surfaces of the connector.
5. Mix the disinfection solution compatible with your cable (see
Solutions Compatibility Table” on page 278
) according to label instructions for solution
strength.
6. Wipe or spray the cable, strain relief, and connector with the disinfectant, following
disinfectant label instructions for wipe durations, solution strengths, and duration of
disinfectant contact with the cable. Ensure that the solution strength and duration of
contact are appropriate for the intended clinical use of the device. Ensure that the
disinfectant solution does not enter the device or the connector or come into contact with
the connector label.
7. Air dry or towel dry with a sterile cloth according to the instructions on the disinfectant
label.
8. Examine the device and cable for damage such as cracks, splitting, sharp edges, or
projections. If damage is evident, discontinue use of the device and contact your Philips
representative.
Transducer Care
Transducer, Cable, and Connector Cleaning
262
EPIQ 7 User Manual 4535 617 25341
Summary of Contents for epiq 7
Page 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Page 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Page 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Page 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Page 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Page 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...