CAUTION
When handling a transducer, do not bump the transducer on hard surfaces.
CAUTION
Do not use a surgeon’s brush when cleaning transducers. Even the use of soft brushes can
damage transducers.
CAUTION
Do not use a brush on the connector label.
CAUTION
During cleaning, disinfection, and sterilization, orient the parts of the transducer, connector,
or cable that must remain dry higher than the wet parts until all parts are dry. This will help
keep liquid from entering unsealed areas of the transducer.
CAUTION
Do not allow any type of fluid to enter the connector. Ensure that fluid does not enter
through the strain relief, through the connector, through the electrical contacts, or through
the areas surrounding the locking-lever shaft and the strain relief. Fluid in the connector may
void the device warranty.
CAUTION
You can use an alcohol solution for disinfection on the connector only. Ensure the solution is
only 70% alcohol or less. Solutions of more than 70% alcohol can cause product damage.
CAUTION
Do not use any alcohol or alcohol-based products on the cable or strain relief.
Transducer, Cable, and Connector Cleaning
Transducer Care
EPIQ 7 User Manual 4535 617 25341
259
Summary of Contents for epiq 7
Page 4: ...4 EPIQ 7 User Manual 4535 617 25341 ...
Page 26: ...Read This First Recycling Reuse and Disposal 26 EPIQ 7 User Manual 4535 617 25341 ...
Page 94: ...DVD RW Drive System Overview System Components 94 EPIQ 7 User Manual 4535 617 25341 ...
Page 154: ...Customizing the System Custom Procedures 154 EPIQ 7 User Manual 4535 617 25341 ...
Page 172: ...Performing an Exam Ending an Exam 172 EPIQ 7 User Manual 4535 617 25341 ...
Page 298: ...System Maintenance For Assistance 298 EPIQ 7 User Manual 4535 617 25341 ...