Installing lamps
Lamp compartment
The lamp compartment is located on the left-hand side of the spectrometer,
beneath a hinged cover. The compartment is easily accessible from the front when
the cover is open.
The spectrometer is designed to operate with PerkinElmer Lumina hollow
cathode lamps (HCLs) and coded electrodeless discharge lamps (EDLs).
The lamp compartment has holders for up to eight lamps.
You can install Lumina HCLs in all eight holders. The holders incorporate the
receptacles for connection of the Lumina lamps.
You can install EDLs in holders 1–4 only. The sockets for the power supplies to
the EDLs are located on the left-hand side of the lamp compartment.
Lamp compartment cover