System Maintenance
0993-5255
4
-11
Replacing the deuterium lamp
The deuterium lamp for the background corrector has a limited lifetime. After a
prolonged period of operation the lamp will exhibit decreasing radiant intensity
and will ultimately fail.
The lifetime of the deuterium lamp will be prolonged if you do not switch it on
and off frequently.
You can replace the deuterium lamp assembly yourself. The replacement lamp is
fitted in a prealigned mount.
Figure 4-2. Location of Deuterium Lamp inside the Spectrometer
1. Switch off the spectrometer at the operational on/off switch on the front panel.
2. Switch the circuit breaker on the rear panel of the spectrometer to the off
position (marked ).
3. Disconnect the spectrometer from the line power supply.
W2.3
Warning: Electrical Hazard
To prevent potential injury to yourself and damage to the instrument, switch
OFF all instruments in the system and disconnect them from the line power
supply before you perform the procedures described below.
Deuterium
lamp cover
Summary of Contents for AAnalyst 700
Page 2: ......
Page 3: ...ATOMIC SPECTROMETRY AAnalyst 700 Atomic Absorption Spectrometer User s Guide...
Page 8: ...Contents C 4 0993 5255...
Page 10: ...1 2 0993 5255...
Page 44: ...Safety Regulatory Information 1 36 0993 5255...
Page 46: ...Spectrometer Installation 2 2 0993 5255...
Page 68: ...Spectrometer Installation 2 24 0993 5255...
Page 70: ...Preparing the System 3 2 0993 5255...
Page 84: ...System Maintenance 4 2 0993 5255...
Page 98: ...Parts and Supplies 5 2 0993 5255...
Page 102: ...Laboratory Requirements 6 2 0993 5255...
Page 120: ...Spectrometer System Description 7 2 0993 5255...
Page 136: ...Translations of Warnings 8 2 0993 5255...
Page 146: ...Translations of Warnings 8 12 0993 5255...
Page 185: ...0993 5255 In 1 Index...
Page 186: ...Index In 2 0993 5255...
Page 192: ......
Page 193: ......