System Maintenance
4
-8
0993-5255
Resetting the automatic circuit breaker
The spectrometer has an automatic circuit breaker which prevents the system
from drawing too much current.
When the circuit breaker is activated, the switch flips to the off position
(marked ). This can happen while an analysis is running.
To reset the switch, wait a few seconds then put the switch to the on position
(marked ).
If this switch is activated frequently, contact your local PerkinElmer office for
advice.
Changing the fuses
1. Switch off the spectrometer and all other instruments in the system as
described in the relevant guides.
2. Switch the circuit breaker on the rear panel of the spectrometer to the off
position (marked ).
3. Disconnect the spectrometer’s line power cord from the line power supply.
4. Disconnect all the line power cords from the sockets at the rear of the
spectrometer.
5. To change the fuses numbered F1, F2, F3, F4:
1. Insert the tip of a suitable screwdriver in the slot of the respective fuse
holder and unscrew the fuse holder.
2. Remove the blown fuse, then insert a new one of the correct type and
rating. The correct rating is shown on the label of the fuse on the rear
panel of the spectrometer.
3. Screw the fuse holder into the correct socket.
6. Reconnect the line power cords correctly.
W2.9
Warning: Electrical Hazard
Use only fuses with the correct current rating and of the specified type for
replacement.
Do not use makeshift fuses and do not short-circuit the fuse holders.
Summary of Contents for AAnalyst 700
Page 2: ......
Page 3: ...ATOMIC SPECTROMETRY AAnalyst 700 Atomic Absorption Spectrometer User s Guide...
Page 8: ...Contents C 4 0993 5255...
Page 10: ...1 2 0993 5255...
Page 44: ...Safety Regulatory Information 1 36 0993 5255...
Page 46: ...Spectrometer Installation 2 2 0993 5255...
Page 68: ...Spectrometer Installation 2 24 0993 5255...
Page 70: ...Preparing the System 3 2 0993 5255...
Page 84: ...System Maintenance 4 2 0993 5255...
Page 98: ...Parts and Supplies 5 2 0993 5255...
Page 102: ...Laboratory Requirements 6 2 0993 5255...
Page 120: ...Spectrometer System Description 7 2 0993 5255...
Page 136: ...Translations of Warnings 8 2 0993 5255...
Page 146: ...Translations of Warnings 8 12 0993 5255...
Page 185: ...0993 5255 In 1 Index...
Page 186: ...Index In 2 0993 5255...
Page 192: ......
Page 193: ......