Laboratory Requirements
6
-14
0993-5255
Furnace gases
Inert gas
The graphite furnace requires a supply of inert gas to prevent the tube and the
analyte atoms from being oxidized when the tube is heated.
Argon is the recommended inert gas.
If you cannot obtain argon, you can use purified nitrogen, but nitrogen does not
provide such efficient protection as argon at temperatures above 2000 °C.
Nitrogen can also react with the samples and the graphite tube at typical
atomization temperatures.
Special gas
For some applications you may want to use a special gas. A typical example is the
use of air during one of the pyrolysis steps to better ash organic materials.
W6.1
Warning: Toxic Gases
If you use nitrogen as the inert gas, toxic cyanogen gas
(CN)
2
will be produced
when the furnace temperature is above 2300 °C.
Make sure that the main laboratory fume ventilation system removes all
gases and fumes emitted by the furnace.
Summary of Contents for AAnalyst 700
Page 2: ......
Page 3: ...ATOMIC SPECTROMETRY AAnalyst 700 Atomic Absorption Spectrometer User s Guide...
Page 8: ...Contents C 4 0993 5255...
Page 10: ...1 2 0993 5255...
Page 44: ...Safety Regulatory Information 1 36 0993 5255...
Page 46: ...Spectrometer Installation 2 2 0993 5255...
Page 68: ...Spectrometer Installation 2 24 0993 5255...
Page 70: ...Preparing the System 3 2 0993 5255...
Page 84: ...System Maintenance 4 2 0993 5255...
Page 98: ...Parts and Supplies 5 2 0993 5255...
Page 102: ...Laboratory Requirements 6 2 0993 5255...
Page 120: ...Spectrometer System Description 7 2 0993 5255...
Page 136: ...Translations of Warnings 8 2 0993 5255...
Page 146: ...Translations of Warnings 8 12 0993 5255...
Page 185: ...0993 5255 In 1 Index...
Page 186: ...Index In 2 0993 5255...
Page 192: ......
Page 193: ......