Laboratory Requirements
6
-16
0993-5255
Cooling the spectrometer system
The graphite furnace uses water cooling to allow the graphite tube to return to its
equilibrium temperature (approximately ambient) quickly between
measurements (
see ‘Cooling the furnace’ in the Graphite Furnace User’s Guide
).
Cooling coils are also incorporated in the spectrometer’s power supply to
maintain the electronic components at a comfortable working temperature.
Cooling system
PerkinElmer offers a recirculating cooling system as an option. It provides
coolant at a constant temperature and flow rate, to allow the most reproducible
analytical conditions. It also makes the system independent of external water
supplies and saves water (
see ‘Cooling the furnace’
in the Graphite Furnace
User’s Guide
). This cooling system includes a pressure-sensitive safety device
that opens a bypass if the internal pressure in the circulatory system increases to
above a safe value.
Power requirements: 230 V; 50/60 Hz (electrical power is supplied by the
spectrometer)
Part Number: B019-8072
External cooling
If you intend to run the spectrometer system without the PerkinElmer cooling
system, you must connect another suitable source of cooling, such as the main
water supply or a laboratory thermostat. If you connect the spectrometer system
to the main water supply, make sure that the water pressure is as given in Table
6-3.
Table 6-3. Requirements for the Cooling Water Supply
Water quality
sediment-free drinking water
Temperature
20–40 °C
Flow rate
not less than 2 L/min
Pressure
2.5–3.5 bar (36–50 psig)
pH
6.5–7.5
Hardness
not greater than 14 °dH or 100 ppm
Summary of Contents for AAnalyst 700
Page 2: ......
Page 3: ...ATOMIC SPECTROMETRY AAnalyst 700 Atomic Absorption Spectrometer User s Guide...
Page 8: ...Contents C 4 0993 5255...
Page 10: ...1 2 0993 5255...
Page 44: ...Safety Regulatory Information 1 36 0993 5255...
Page 46: ...Spectrometer Installation 2 2 0993 5255...
Page 68: ...Spectrometer Installation 2 24 0993 5255...
Page 70: ...Preparing the System 3 2 0993 5255...
Page 84: ...System Maintenance 4 2 0993 5255...
Page 98: ...Parts and Supplies 5 2 0993 5255...
Page 102: ...Laboratory Requirements 6 2 0993 5255...
Page 120: ...Spectrometer System Description 7 2 0993 5255...
Page 136: ...Translations of Warnings 8 2 0993 5255...
Page 146: ...Translations of Warnings 8 12 0993 5255...
Page 185: ...0993 5255 In 1 Index...
Page 186: ...Index In 2 0993 5255...
Page 192: ......
Page 193: ......