Spectrometer Installation
0993-5255
2
-13
To connect both argon and a special gas
Refer to Figure 2-3 on the following page when you connect the furnace gases.
1. Make sure that the gas supplies are turned off.
2. Use the special tube cutter to cut the special gas tube into two pieces for the
two gas supplies. The ends of the tubes must be cut squarely.
3. Connect the gas tubes to the argon (or nitrogen) and special gas pressure
regulators. Use the R ¼'' screw connectors provided or use a connector or
adapter suitable for the gas regulator. Make sure that all the connections are
secure.
4. Pass the gas tubes to the rear of the spectrometer. Make sure that they cannot
be damaged or strained.
5. Push the free ends of the gas supply tubes into the
Normal Gas
and
Special
Gas
connectors on the manifold as shown.
Push the tubes into the connectors until the tubes butt up to the end of the
connector.
Note:
To disconnect the gas tube from a connector:
Carefully push the black locking ring of the connector into the connector and
pull the tube out of the connector.
Summary of Contents for AAnalyst 700
Page 2: ......
Page 3: ...ATOMIC SPECTROMETRY AAnalyst 700 Atomic Absorption Spectrometer User s Guide...
Page 8: ...Contents C 4 0993 5255...
Page 10: ...1 2 0993 5255...
Page 44: ...Safety Regulatory Information 1 36 0993 5255...
Page 46: ...Spectrometer Installation 2 2 0993 5255...
Page 68: ...Spectrometer Installation 2 24 0993 5255...
Page 70: ...Preparing the System 3 2 0993 5255...
Page 84: ...System Maintenance 4 2 0993 5255...
Page 98: ...Parts and Supplies 5 2 0993 5255...
Page 102: ...Laboratory Requirements 6 2 0993 5255...
Page 120: ...Spectrometer System Description 7 2 0993 5255...
Page 136: ...Translations of Warnings 8 2 0993 5255...
Page 146: ...Translations of Warnings 8 12 0993 5255...
Page 185: ...0993 5255 In 1 Index...
Page 186: ...Index In 2 0993 5255...
Page 192: ......
Page 193: ......