Safety/Regulatory Information
0993-5255
1
-23
Hazards with electrothermal atomization
Toxic products:
Toxic metal vapors, ozone, etc., can be generated by the system, depending on the
type of analyses being performed.
You must provide an efficient laboratory ventilation system to remove toxic
products generated during instrument operation.
If you use nitrogen as the inert gas, toxic cyanogen gas (CN)
2
will be produced
when the furnace temperature is above 2300 °C.
High temperatures:
The HGA graphite furnace can generate temperatures of up to 3000 °C
To avoid serious burns, never touch the graphite components until the furnace
has cooled to room temperature.
Do not attempt to inject a sample into the graphite tube when the furnace is
hot. You may be subjected to fumes from the sudden vaporization of the
sample, and may damage the pipet.
UV radiation:
The graphite furnace emits ultraviolet radiation when it is heated to
incandescence.
Never directly view the graphite furnace unless you are wearing UV-
absorbing glasses.
Summary of Contents for AAnalyst 700
Page 2: ......
Page 3: ...ATOMIC SPECTROMETRY AAnalyst 700 Atomic Absorption Spectrometer User s Guide...
Page 8: ...Contents C 4 0993 5255...
Page 10: ...1 2 0993 5255...
Page 44: ...Safety Regulatory Information 1 36 0993 5255...
Page 46: ...Spectrometer Installation 2 2 0993 5255...
Page 68: ...Spectrometer Installation 2 24 0993 5255...
Page 70: ...Preparing the System 3 2 0993 5255...
Page 84: ...System Maintenance 4 2 0993 5255...
Page 98: ...Parts and Supplies 5 2 0993 5255...
Page 102: ...Laboratory Requirements 6 2 0993 5255...
Page 120: ...Spectrometer System Description 7 2 0993 5255...
Page 136: ...Translations of Warnings 8 2 0993 5255...
Page 146: ...Translations of Warnings 8 12 0993 5255...
Page 185: ...0993 5255 In 1 Index...
Page 186: ...Index In 2 0993 5255...
Page 192: ......
Page 193: ......