X - R i t e e X a c t 2
63
Technical Specifications
SPECTRAL ENGINE
Spectral analyzer:
DRS spectral engine
Spectral range:
400 nm - 700 nm with 10 nm interval
OPTICS
Measurement geometry
:
45°:0° ring illumination optics, ISO 13655:2017
Measurement aperture
:
1.5 mm, 2 mm, 4 mm or 6 mm
Light source:
LED (Full spectrum LED light source)
REFLECTANCE MEASUREMENT
Measurement Conditions:
According to ISO 13655:2017:
• M0: UV included
• M1: (method 2) D50
• M2: UV excluded
• M3: Polarization (not in eXact 2 Xp).
Calibration:
Automatic on white reference
Inter instrument agreement:
Average:0.25
Δ
E*ab, Max: 0.45
Δ
E*ab (M3 0.55
Δ
E*ab)
(Measurements using X-Rite manufacturing standards at a
temperature of 23°C +/- 1°C, 40-60% RH for all measurement
modes on 12 BCRA color tiles and a white ceramic reference
(D50, 2°))
Short term repeatability
White
: 0.02
Δ
E*ab, (standard deviation) White BCRA
(Error compared to mean value of 20 measurements every 5
seconds)
Short term repeatability
Density:
+/-0.01 D for CMYK measurements
Scan Length:
Max: 1,120 mm (44”)
OPERATING CONDITIONS
Operating Temperature Range:
50° to 95°F (10° to 35°C)
30-85% relative humidity maximum (non-condensing)
Storage Temperature Range
:
-4° to 122°F (-20° to 50°C)
INTERFACE, DIMENSIONS AND WEIGHT
Data Interface:
USB
USB-C port (delivered with adapter to connect to USB-A ports)
Wi-Fi
Supports IEEE 802.11 b/g/n standards
Supports 2.4 GHz ISM band only (not 5 GHz).
Summary of Contents for X-Rite eXact 2
Page 1: ...User Guide ...
Page 2: ......