Chapter 4 Explanation of Functions
4-78
4
Timing Mode
・
When buffering operation starts, the timing input is turned to stand-by status.
・
Changing the timing input from “ON” to “OFF” during stand-by status starts data
accumulation to the controller memory.
・
Timing input stops when the accumulation amount has reached the setting value
or when buffering operation stops.
Changing the timing input from “ON” to “OFF” again deletes the previous data
and starts new data accumulation. Make sure to stop the buffering operation once
and load/save the accumulated measurement data before performing the next
accumulation.
Sample Trigger Mode
· When buffering operation starts, the trigger generation is turned to stand-by status.
· Accumulation of the measurement data for the setting sample trigger accumulation
amount starts after the setting trigger condition is generated.
· After completion of sample trigger accumulation, the trigger generation is turned to
stand-by status again.
· When the setting trigger condition is generated again, accumulation of the setting
sample trigger amount starts.
· Accumulation operation stops when the accumulation amount has reached the
setting value or when buffering operation stops.
Be sure to set the sample trigger accumulation amount so that (accumulation
amount) ÷ (sample trigger accumulation amount) is an integer value.
Timing input ON accumulation
amount reached
Stand-by
Timing input
OFF
Timing input
ON
OFF
Accumulation of data
Meas ope
Buffering start
Buffering stop
Stand-by
Trigger
generate
Meas ope
Buffering start
Buffering stop or
Accu. amount reached
Trigger
generate
Accumulation of data
Stand-by
Accumulation of data
Stand-by
Sample trig. accu.
amount reached
Summary of Contents for HL-C2 Series
Page 32: ...Preface 30 MEMO ...
Page 50: ...Chapter 1 Prior to Use 1 18 1 MEMO ...
Page 65: ...Chapter 3 Programmable Display Operation 3 5 3 MEMO ...
Page 164: ...Chapter 4 Explanation of Functions 4 90 4 MEMO ...
Page 191: ...6 1 6 6 Maintenance and Inspection 6 1 Maintenance 6 2 6 2 Inspection 6 2 ...
Page 200: ...Chapter 7 Troubleshooting 7 8 7 MEMO ...
Page 250: ...Chapter 8 Specifications 8 50 8 50mm type Unit mm 85mm type Unit mm ...
Page 252: ...Chapter 8 Specifications 8 52 8 8mm type Unit mm 15mm type Unit mm ...
Page 270: ...Chapter 8 Specifications 8 70 8 MEMO ...
Page 271: ...1 Apx Appendix 1 Index 2 ...
Page 275: ...Appendix 5 Apx MEMO ...
Page 277: ......