Chapter 4 Explanation of Functions
4-70
4
■RS-232C Output
This function is used for performance setting of the measurement data
output from RS-232C to an external device.
● RS-232C Output mode
RS-232C output is selectable from three modes.
The initial value is set to “Handshake”.
Setting item
Contents
Handshake
In response to a request command from an upper device,
transmits the result data as a reply command.
Timing
Transmits the measured value in the exclusive output format at
timing input ON during Timing mode is selected.
Continuous
Transmits the measured value continuously in the exclusive
output format after Continuous mode is selected.
Refer to the page 5-12 for details of the exclusive output format.
Note that the command send/receive when RS-232C output mode is set to
Continuous
or
Timing
operates as the followings:
[RS-232C port]
Any other commands than “RS-232C Output Mode” is ignored. When
“RS-232C Output Mode” is received, it transmits the reply and shifts to
each mode.
[USB port]
Send/receive of all commands is available.
● RS-232C output type
When
Timing
or
Continuous
is selected for “RS-232C Output Mode”, an
output destination of the measurement data is selectable from OUT1, OUT2,
and both OUT1&OUT2.
The initial value is set to
“OUT1”
.
Summary of Contents for HL-C2 Series
Page 32: ...Preface 30 MEMO ...
Page 50: ...Chapter 1 Prior to Use 1 18 1 MEMO ...
Page 65: ...Chapter 3 Programmable Display Operation 3 5 3 MEMO ...
Page 164: ...Chapter 4 Explanation of Functions 4 90 4 MEMO ...
Page 191: ...6 1 6 6 Maintenance and Inspection 6 1 Maintenance 6 2 6 2 Inspection 6 2 ...
Page 200: ...Chapter 7 Troubleshooting 7 8 7 MEMO ...
Page 250: ...Chapter 8 Specifications 8 50 8 50mm type Unit mm 85mm type Unit mm ...
Page 252: ...Chapter 8 Specifications 8 52 8 8mm type Unit mm 15mm type Unit mm ...
Page 270: ...Chapter 8 Specifications 8 70 8 MEMO ...
Page 271: ...1 Apx Appendix 1 Index 2 ...
Page 275: ...Appendix 5 Apx MEMO ...
Page 277: ......