Chapter 4 Explanation of Functions
4-48
4
■ Analog Output at Alarm
This function sets the analog output status when an alarm is issued.
It can select whether holding the previous value or setting a fixed value for the
analog output under an alarm operation status (poor light intensity, or measurement
disabled when a target object goes off the measurement range).
· The input range of fixed value is -10.800 to +10.800[V], and the initial value is
+10.800[V].
· The current output varies on load impedance. The maximum value is 25.0[mA].
· Check the input range of AD board before setting.
· Before setting of analog output at alarm, read
“4-4-1 Alarm Setting and
Outputs
” carefully.
● How to use
1
Display the “Analog Output at Alarm” screen.
2
Touch the “Analog Output at Alarm” key.
The display is changed between
Hold Pre Val
and
Fixed Value
each
time the key is touched.
When selecting “Fixed Value”
3
Touch the
key.
The screen for inputting the fixed value appears.
4
Input the fixed value.
(1) Touch the
OUT1
(or
OUT2
) key to
prepare for inputting the setting value.
(2) Input the value.
(3) Touch the
key to confirm the value.
Selection screen of OUT1
Selection screen of OUT2
2
3
1
Summary of Contents for HL-C2 Series
Page 32: ...Preface 30 MEMO ...
Page 50: ...Chapter 1 Prior to Use 1 18 1 MEMO ...
Page 65: ...Chapter 3 Programmable Display Operation 3 5 3 MEMO ...
Page 164: ...Chapter 4 Explanation of Functions 4 90 4 MEMO ...
Page 191: ...6 1 6 6 Maintenance and Inspection 6 1 Maintenance 6 2 6 2 Inspection 6 2 ...
Page 200: ...Chapter 7 Troubleshooting 7 8 7 MEMO ...
Page 250: ...Chapter 8 Specifications 8 50 8 50mm type Unit mm 85mm type Unit mm ...
Page 252: ...Chapter 8 Specifications 8 52 8 8mm type Unit mm 15mm type Unit mm ...
Page 270: ...Chapter 8 Specifications 8 70 8 MEMO ...
Page 271: ...1 Apx Appendix 1 Index 2 ...
Page 275: ...Appendix 5 Apx MEMO ...
Page 277: ......