11.2.2
Reflex screening
In the
Tympanometry
test screen, the Reflex Screening sec-
tion shows reflex screening values if reflex screening meas-
urements are available, or crossed out values if they are not.
If, subsequently, a Reflex Threshold test (see
) is made using the same stim-
ulus types as in Reflex Screening, the results from the Reflex
Threshold test will replace the Reflex Screening values
shown in this view. Any stimulus types not tested during a
Reflex Threshold test will remain in the Reflex Screening res-
ults table.
If sequence testing is selected, the test starts with tympanometry testing and continues with reflex screening.
Procedure
If sequence testing is selected, the test starts with tympanometry testing and continues with reflex screening.
1. In order to prepare MADSEN OTOflex 100 to start measuring as soon as seal is achieved, do the following:
2. Prepare the patient for testing. See
3. Load a test setting configured for screening patients in the relevant age group. See recommended settings below.
Note
•
The safety intensity level cannot be exceeded in screening mode.
4. If needed, toggle the
Ear Selector
to select the ear on which you wish to start the test.
Sequence testing
If sequence testing is selected, the test starts with tympanometry testing and continues with
reflex screening.
–
In the
Tympanometry
screen, select
T+RS
in the
Sequence
field below the Control
Panel.
–
Without applying the eartip to the patient’s ear click
Start
in the
Sequence
section of the
Control Panel.
Reflex screening only
–
Without applying the eartip to the patient’s ear click
Start
in the
Sequence
section of the
Control Panel.