14.2.4
ETF-P view settings
Time axis
Scale(sec.)
When the end of the time axis is reached, it will automatically
extend with 10 more seconds up to the max time value, 60 seconds.
The graph can be rescaled to show more detail, even after the meas-
urement has been completed.
Default value: 30 seconds.
14.3
Measurement settings
Measurement settings are your way of defining how the test results should be carried out.
14.3.1
Tympanometry setup
When you perform a tympanometry sweep, there are a number of settings you can adjust. These settings are available
either in the Control Panel or in the
Tools
>
Options
>
Tymp. and Reflex Scr.
dialog.
14.3.1.1
Tympanometry: Control Panel settings
Tympanometry
Probe Tone
Select the frequency for this measurement setup.
226 Hz
Adult default
If 226 Hz is selected, you can set admittance to be shown in the
admittance unit mmho or in the compliance volume units ml, cc or
cm
3
as described in
Tymp. and Reflex Scr. view settings
.
1000 Hz
Infant default
Recommended for testing on infants younger than 4-6 months. See
also the MADSEN OTOflex 100 Guide.
For frequencies higher than 226 Hz, admittance is shown
only
in
mmho. The value in
Y scale
automatically changes to mmho.
Pump speed
Changes the speed of the pressure sweep. Indicated in daPa per
second.
•
50, 100, 200, 400 daPa/s, or
•
AFAP
(As Fast As Possible, 500-600 daPa/s in 2 - 0.5 cc): Forces
the pump to work as fast as possible. The actual pressure build-up
depends on the amount of air escaping from probe and ear.
Default:
AFAP
daPa/s
Pressure range
You can choose between
Normal range
and
Custom range
:
150