Pressure build functions
Softkeys
Pressure build functions depending on the setting of
Initial pressure
.
Right
Pressure control
Activate
Pressure Up
or
Pressure Down
to automatically build up pressure until
the first opening of the Eustachian tube or max. pressure is reached. The pressure
control is automatically reversed after one of the two criteria has been reached.
Right
Pressure control
Press to automatically build up negative pressure until the first opening of the Eus-
tachian tube or max. pressure is reached. The pressure control is automatically
reversed after one of the two criteria has been reached.
Left
Stop
and go to idle mode.
Press and hold to display patient and user info.
Scroll wheel
Optional method: Turn the scroll wheel clockwise (positive) or counter-clockwise
(negative) to manually control the pump instead of using the pressure control
softkeys.
The manual pressure control works in accordance with the current softkey symbol.
8.10.2
ETF-P - procedure
A.
Scroll wheel
Manual pressure build-up
B.
Start test
/
Automatic pressure build-up
/
Procedure
1. Press
On
/
Off
briefly to select the
Tymp
test screen.
2. Record a tympanogram to confirm the perforation. The resulting tym-
panogram curve will be flat, and the ear canal volume will be abnormally
high because it includes the volume of the entire middle ear.