Procedure
Prepare MADSEN OTOflex 100 to start measuring as soon as seal is achieved:
1. Do as described in
.
2. Load a test setting configured for screening patients in the relevant age group. See recommended settings below.
Note
•
The safety intensity level cannot be exceeded in screening mode.
3. If needed, toggle the
Ear Selector
to select the ear on which you wish to start the test.
Sequence testing
If sequence testing is selected, the test starts with tympanometry testing and continues with
reflex screening.
–
Select
Menu
>
Procedure options..
>
T+RS
.
–
Without applying the eartip to the patient’s ear press and hold to start the sequence.
Reflex screening only
–
Without applying the eartip to the patient’s ear press to start the test.
4. Apply the probe to the patient’s ear with a steady grip. This will prevent probe movements
from resulting in leakage or changes in admittance.
The test starts automatically as soon as a good probe fit with an airtight seal is achieved.
Warning
•
If the patient is troubled by the test, stop the test. The test is interrupted and
the pump pressure is relieved immediately.Already measured results are kept.
or
Pausing the test
–
Steady: Press to pause test.
–
Flashing: Press to continue.
After the pause, the test resumes at the intensity it came to or was interrupted in.
Stopping the test
–
Press the
Stop
softkey to stop test.
Recommended settings Reflex screening
Setting
Value
Auto start on seal
(
Menu
>
Procedure options..
>)
On
64
Otometrics - MADSEN OTOflex 100
8 Testing with MADSEN OTOflex 100