13.6
Acoustic Reflex Decay setup
The following settings are useful:
Probe Tone
Menu
>
Decay can be measured only with a 226 Hz probe tone.
Default: 226 Hz
With a probe tone at 226 Hz, you can set admittance to be shown in the admittance unit mmho or
in the compliance volume units ml, cc, cm
3
or
µ
l.
Use TPP offset
Menu
>
Select or deselect the use of pressure offset to stabilize the eardrum. To change the current set-
ting, select
Menu
>
More settings..
>
TPP offset
.
Off
: TPP is used if available
On
: TPP + TPP offset (according to TPP sign) are used
Default value:
On
TPP offset
Menu
>
More settings..
Select a pressure offset to stabilize the eardrum. This may facilitate reflex recordings from ears
showing steeply sloped tympanograms. A positive value offsets in same direction as TPP sign.
-50 < x < 50 daPa (offset according to TPP sign)
Default value: 20 daPa
Pump during
stim.
Menu
>
More settings..
On
/
Off
If
On
, the pump will actively keep the TPP pressure even during stimulus presentation.
If
Off
, the pump will adjust the pressure between presentations.
Default value:
Off
Deflection scale
Menu
>
More settings..
Enter the appropriate setting for the reflex decay deflection scale.
The applicable unit (mmho, cc, ml, cm
3
or
µ
l) follows unit and probe tone settings.
If
µ
l is used for 226 Hz probe tone, then all values are multiplied by 1000.
If a negative scale is selected, the graph by default leaves room for negative deflections and vice
versa. However, the zero axis will automatically move if the deflection graph should go in the
opposite direction.
Default: -0.30 mmho
Show 500 Hz
Menu
>
More settings..
Select the stimulus type to be included in reflex decay testing.
Default:
On
142
Otometrics - MADSEN OTOflex 100
13 Test settings in MADSEN OTOflex 100