Procedure
5. Apply the probe to the patient’s ear with a steady grip. This will prevent probe movements
from resulting in leakage or changes in admittance.
The test starts automatically as soon as a good probe fit with an airtight seal is achieved.
If not, press
Start
on the Control Panel to prepare to start measuring as soon as seal is
achieved.
Warning
•
If the patient is troubled by the test, stop the test. The test is interrupted and
the pump pressure is relieved immediately.Already measured results are kept.
or
Pausing the test
–
Steady: Press to pause test.
–
Flashing: Press to continue.
After the pause, the test resumes at the intensity it came to or was interrupted in.
Stopping the test
–
Press to stop test.
Recommended settings - Reflex screening
Setting
Value
Toolbar icon
Pump speed
(Control Panel)
AFAP
(As Fast As Possible).
Auto start on seal
(
Tools
>
Options
>
General
> )
If set to
Sequence
:
The test starts automatically as soon as a good probe fit with an airtight
seal is achieved.
Sequence
Auto resume on seal
(
Menu
>
Procedure options..
>)
This function is ideal for speeding up the test procedure.
–
On
:
If the screening eartip is removed from the patient’s ear, the
test continues from the intensity it came to or was interrupted
in, when there is seal.
–
Off
:
The screening eartip must remain in place against the patient’s
ear. If the probe is removed, the test is interrupted.
On
Otometrics - MADSEN OTOflex 100
95
11 Testing with the OTOsuite Immittance module