T + RT + RD
(Tympan Reflex Thr Reflex Decay)
–
Off
(MADSEN OTOflex 100)
The press-and-hold function is disabled, and only the currently selected measurement type will be performed.
8.3.2
Running a test sequence from MADSEN OTOflex 100
In test mode the sequence currently selected is shown at the top of the test screen.
Starting a sequence
•
When you have selected a sequence, press and hold the right softkey to start the sequence.
Note
•
The sequence starts as soon as probe seal is achieved.
Menu
>
Procedure options..
>
Auto start on seal
must be set to
Sequence
.
Warning
•
If the patient is troubled by the test, stop the test. The test is interrupted and the
pump pressure is relieved immediately.Already measured results are kept.
Stopping a sequence
If a leak is detected or the probe is removed from the patient's ear, the sequence pauses automatically.
You can stop the sequence manually, for instance by pressing the
Pressure Release
key on MADSEN
OTOflex 100.
Resuming a sequence
•
If
Menu
>
Procedure options..
>
Auto resume on seal
is
On
, and a leak is detected or the probe is removed from the
patient's ear, the sequence pauses automatically. The sequence will automatically resume from the pause state when
seal has been reestablished.
The sequence continues from where it was paused. An interrupted reflex stimulus will be repeated.
Note
•
If the leak occurs during the tympanometric measurement, the tympanometry part of the sequence will
not be repeated. Press
Stop
and restart the sequence if you need to repeat the tympanometric measurement.
8.4
Screening
You can perform Tympan Reflex Screening (T+RS) as a sequence from the Tympanometry test screen.
Note
•
The safety intensity level cannot be exceeded in screening mode.
Note
•
You can use either ordinary eartips or screening eartips for screening.
If you are using screening eartips and the setting
Auto start on seal
is set to
On
, the test will start with
Otometrics - MADSEN OTOflex 100
61
8 Testing with MADSEN OTOflex 100