14.3.5.1
Reflex Decay: Control Panel settings
Acoustic reflex decay is tested using a 226 Hz probe tone. It is typically not recommended to perform supra-threshold
reflex testing, such as decay, on infants.
Reflex Decay
Custom TPP offset
Select or deselect the use of pressure offset to stabilize the eardrum.
To change the current setting,, if
Control Panel
>
Custom TPP offset
is checked, select
Tools
>
Options
>
Reflex Decay
>
Measurement
>
General
>
Custom TPP offset
.
Disabled
: TPP is automatically set to
25 daPa more
.
Enabled
: TPP + TPP offset (according to TPP setting in
Tools
>
Options
>
Reflex Decay
>
Measurement
>
Custom TPP offset
)
are used.
Default value:
Enabled
Manual
Stimulus
Select the stimulus types to be included.
Available types: 500 Hz, 1000 Hz.
Adult default: 1000 Hz
Infant default: Not currently available
Intensity
Select the stimulus intensity from the drop-down list.
If no reflex threshold is available, the
Control Panel
>
Intensity
set-
ting is
Off
. This is useful for admittance monitoring.
Default: Thr10 dB, or
Off
•
Stimulus levels at or above the warning level
(> 108 dB SPL re 2cc) are marked by an orange bar in the
Intens-
ity
drop-down list.
•
Stimulus levels at or above the intensity safety level
(> 115 dB SPL re 2cc) are marked by a red bar in the
Intensity
drop-down list.
Note
•
The
Intensity increment
used in Reflex Decay is set in
the Reflex Threshold setup. This is because the Reflex Decay
stimulus level defaults to the reflex thr10 dB.
Stimulus side
Adult default: Ipsi
Infant default: Ipsi
•
Ipsi
Stimulus is presented in the probe ear where the measurements
are made.
•
Contra
Stimulus is presented through an earphone or insert phone in
the ear opposite to the probe ear.