Note
•
Wipe the plastic probe tip with a disinfectant (for example ethanol) between patients or replace it with a
spare one.
•
If you have replaced the probe tip and/or the acoustic filter, do a probe test.
NEVER
insert the probe tip into the test
cavity of MADSEN OTOflex 100 without first cleaning and disinfecting the probe tip. See
15.4
Cleaning and disinfecting the probe tip
Caution
•
Thorough cleaning of the probe tip is required after use in infected ear canals. Cleaning the threaded
ring may also be necessary.
Note
•
Always comply with local hygienic standards for disinfection.
1. To remove the probe tip, hold
the probe by the probe body
and unscrew the threaded ring.
Take out the probe tip.
A.
Probe body
B.
Threaded ring
C.
Probe tip
2. Check to see if the sound chan-
nels of the probe tip are
blocked. If they are, use the
cleaning wire to clean the
sound channels.
Always clean from the rear.