Probe Tone
Menu
>
Scroll to select the appropriate frequency for this measurement setup.
Adult default: 226 Hz
Infant default: 1000 Hz
If 226 Hz is selected, you can set admittance to be shown in the admittance unit mmho or
in the compliance volume units ml, cc or cm
3
as described in
.
For frequencies higher than 226 Hz, admittance is shown
only
in mmho. The value on the
Admittance scale
automatically changes to mmho.
•
226 Hz
•
1000 Hz
Recommended for testing on infants younger than 4-6 months. See also
Baseline comp(ensa-
tion)
Menu
>
On
:
The middle ear admittance alone (after subtraction of the ear canal contribution) is meas-
ured.
Off
:
The total ear admittance (including the ear canal) is measured.
Default:
On
Tymp scale
Menu >
This setting follows unit, probe tone, and baseline comp. settings.
Default: 1.5 mmho 226 Hz
Default: 3.0 mmho 1000 Hz
Layered curves
Menu
>
When
Layered curves
is
On
, this affects a number of areas:
•
If
Auto Scale
is
checked
, the graph will be scaled to allow for showing the highest
curve. Auto scaling is still ear dependent.
•
If a 1000 Hz tympanogram is included among the layered curves, the applicable units
and tymp results labels for 1000 Hz probe tone measurements will also be used for the
226 Hz curves.
•
The widest pressure range used will be applied.
The selected curve is highlighted and shown as a bold full line.
Note
•
This setting is not saved as part of a measurement setup.
Default:
On
132