Procedure
4. Apply the probe to the patient’s ear with a steady grip. The test starts automatically as soon as
a good probe fit with an airtight seal is achieved.
Warning
•
If the patient is troubled by the test, stop the test. The test is interrupted and
the pump pressure is relieved immediately.Already measured results are kept.
or
Pausing the test
–
Steady: Press to pause test.
–
Flashing: Press to continue.
Stopping the test
–
Press to stop test.
Recommended settings
Setting
Value
Pump speed
(
Menu
>
More settings..
>)
AFAP
(As Fast As
Possible).
Auto start on seal
(
Menu
>
Procedure options..
>)
On
Auto resume on seal
(
Menu
>
Procedure options..
>)
This function is ideal for speeding up the test procedure.
On
Stop when results are available
(
Menu
>
More settings..
>)
The measurement stops automatically when satisfactory values have been registered.
On
8.4.2
Reflex screening
A.
Press to start the test, if
Auto start on seal
is set to
Off
If you use screening eartips, it is recommended that you use the setting
Menu
>
Pro-
cedure options..
>
Auto start on seal
set to
Sequence
. When a good probe fit with
an airtight seal is achieved, i.e. when the screening eartip is pressed gently against
the ear, testing will start with a tympanometric test and continue with this test.
Otometrics - MADSEN OTOflex 100
63
8 Testing with MADSEN OTOflex 100