![otometrics MADSEN OTOflex 100 Reference Manual Download Page 130](http://html.mh-extra.com/html/otometrics/madsen-otoflex-100/madsen-otoflex-100_reference-manual_1658679130.webp)
Device Info.
Displays technical information about hardware, firmware and Bluetooth.
Brightness
For adjusting the brightness of the display.
Wheel click vol.
For adjusting the wheel click volume.
Button click vol.
For adjusting the button click volume.
Bluetooth
Defines whether Bluetooth communication is on or off. Allows you to completely dis-
able Bluetooth communication in case MADSEN OTOflex 100 is to be used in extremely
radio sensitive environments.
Battery type
NiMH or Alka(line). Must be set according to the battery type inserted in the battery
compartment. If MADSEN OTOflex 100 suspects that the setting is wrong (based on bat-
tery voltage checks), you will be prompted once for selection of the battery type.
Altitude ab. sea
The altitude above sea level affects the barometric pressure and thus the air density.
This setting is used to optimize the pump operation according to the current air density.
The setting is listed in increments of 100 meters (100 meters = approx. 330 ft).
-100 to 4000 metres.
Bat(tery) power save
This setting applies when MADSEN OTOflex 100 is running on battery power. After this
time period in active mode without any user actions on MADSEN OTOflex 100 or via
OTOsuite, MADSEN OTOflex 100 will enter power save mode where the display changes
to a power save screen and the auto start and auto resume functions stop working.
Power save mode reduces power consumption and preserves battery capacity.
Recommended setting: 3 minute or as short as possible.
If the battery capacity is low, MADSEN OTOflex 100 will power off instead of entering
power save.
Press any key or operate MADSEN OTOflex 100 from OTOsuite to resume full operation.
Bat(tery) power off
This setting applies when MADSEN OTOflex 100 is running on battery power. After this
time period in power save mode MADSEN OTOflex 100 will power off. Power off pre-
serves battery capacity.
Recommended setting: 7 minutes or as short as possible.
130
Otometrics - MADSEN OTOflex 100
13 Test settings in MADSEN OTOflex 100