The Tymp test screen
The
Tymp
test screen provides the tympanometric curves,
norm area, the resulting Type, TPP, and ECV/EBV for the curve
shown.
Note
•
The probe tone shown in the test screen applies
to the currently selected measurement setting, and
may not reflect the probe tone setting that was used
for the curve shown.
The actual probe tone applying to the curve shown can be
found in the
Tymp Data
screen (see below).
The tympanometry graph shows a maximum of 3 curves per
ear per patient. They are shown either all at once as layered
curves, or individually.
•
Use the
Curve Selector
) to select a specific curve for viewing.
The results for the highlighted curve are displayed.
•
To return to the standard
Tymp
view, press again.
The Tymp Data screen
1. If needed, press
On
/
Off
briefly to select the
Tymp Data
screen.
The results shown in the
Tymp Data
screen reflect the curve cur-
rently selected in the
Tymp
test screen.
The following features are described in detail in
–
Tympanometric Peak Pressure, TPP
–
Static Admittance, SA (Static Compliance, SC, when volume units
are used)
–
Tympanometric Width, TW
–
Equivalent Ear Canal Volume, ECV (Equivalent Baseline Volume,
EBV, when 1000 Hz probe tone is used)
–
Tympanometry type, “Type”
–
Probe tone, P, in Hz
2. Press
Select
briefly to return to the standard
Tymp
test screen.
Otometrics - MADSEN OTOflex 100
69
8 Testing with MADSEN OTOflex 100