Stimulus Intensities
Max. (dB HL)
When the list is shown, simply click on a specific intensity to select.
Then consider which adjacent intensities you wish to include. The
options will show according to the selected step size.
Note
•
You cannot extend the range to include stimulus levels
exceeding 100 dB HL in this test.
Whenever the stimulus intensity level exceeds 108 dB SPL, for
example when you select 100 dB HL at 500 Hz Ipsi, a warning will
appear, and the automatic test is paused. You will be prompted
whether to continue or skip to the next stimulus type.
Note
•
To avoid automatic testing being interrupted because
of high stimulus intensity levels when reaching the warning
limits, it is recommended that you set the max. intensity to
95 dB HL.
Increment (dB)
Select the step size to increment the stimulus intensity (Sound Pres-
sure Level) in increments of 10, 15, or 20 dB.
Default: 10 dB
Number of intens-
ities
Depending on the increment selected, you can select or deselect the
intensity. You can select up to three stimulus intensities. The com-
plete range of intensities available will appear when you deselect all.
Stimulus side
Stimulus side
•
Ipsi
Stimulus is presented in the probe ear where the measurements
are made.
•
Contra
Stimulus is presented through an ear phone or insert phone in
the ear opposite to the probe ear.
•
Ipsi then Contra
•
Contra then Ipsi
Default: Ipsi
Stimulus types
Stimulus types
Select the stimulus types to be included/excluded. At least one stim-
ulus type must always be selected.
Available types: 500 Hz, 1000 Hz, 2000 Hz, 4000 Hz, BBN.
14.3.4
Reflex Threshold setup
When you perform a Reflex Threshold measurement, there are a number of settings you can adjust. These settings are
available either in the Control Panel on the left side of the screen, or in the
Tools
>
Options
>
Reflex Threshold
>
Meas-
urement
dialog.
154
Otometrics - MADSEN OTOflex 100
14 Immittance module settings