No. of intensities
Menu
>
More settings..
1, 2 or 3. Define whether you wish to use more than one intensity during testing.
If set to 1, testing uses only max intensity.
If set to 2, testing starts at max intensity - intensity increment and continues with max
intensity if no reflex was found.
If set to 3, testing starts at max intensity - 2 * intensity increment and continues up to
max intensity or until a reflex is detected.
Default: 1
Intensity increment
Menu
>
More settings..
Select the step size to increment the stimulus intensity (Sound Pressure Level) in incre-
ments of 10, 15, or 20 dB.
Default: 10 dB
Reflex criterion
Menu
>
More settings..
Enter the appropriate setting for the reflex criterion. It follows unit and probe tone set-
tings.
The applicable unit (mmho, cc, ml, cm
3
or
µ
l) is selected automatically, when “Unit” is
set.
If
µ
l is used for 226 Hz probe tone, then all values are multiplied by 1000.
Default criterion: ±0.02 mmho
13.5
Acoustic Reflex Threshold setup
The following settings are useful:
Probe Tone
Menu
>
Scroll to select the appropriate frequency for this measurement setup.
Adult default: 226 Hz
Infant default: 1000 Hz
With a probe tone at 226 Hz, you can set admittance to be shown in the admittance
unit mmho or in the compliance volume units ml, cc, cm
3
or
µ
l.
For frequencies higher than 226 Hz, admittance is shown
only
in mmho. The value in
Y
scale
automatically changes to mmho.
•
226 Hz
•
1000 Hz
Recommended for testing on infants younger than 4-6 months. See
Otometrics - MADSEN OTOflex 100
137
13 Test settings in MADSEN OTOflex 100